Tosoh Bioscience Inc: Medical Device Recall in 2018 - (Recall #: Z-0600-2019)
See the recall detail below. You can also see other recalls from the same firm in 2018.
AIA-900Analyzer, Product code 022930, 022930R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.
Class II
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Affects all active analyzers. 022930, 022930R
US Nationwide distribution and International distribution to Colombia, Chile, Honduras, Cayman Islands, Panama, Costa Rica, Guatemala, Venezuela, Uruguay, Peru, and Ecuador.
Voluntary: Firm initiated