Tosoh Bioscience Inc: Medical Device Recall in 2020 - (Recall #: Z-2967-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay
Product Classification:
Class II
Date Initiated: August 7, 2020
Date Posted: September 23, 2020
Recall Number: Z-2967-2020
Event ID: 86218
Reason for Recall:
Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
Status: Terminated
Product Quantity: 289 boxes
Code Information:
Lot Numbers: J833531, JX33532, JZ33533, A233534, A433535
Distribution Pattern:
Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.
Voluntary or Mandated:
Voluntary: Firm initiated