Product Description:

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.

Product Classification:

Class II

Date Initiated: July 31, 2015
Date Posted: September 16, 2015
Recall Number: Z-2738-2015
Event ID: 71936
Reason for Recall:

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Status: Terminated
Product Quantity: 108 units
Code Information:

TSL002605

Distribution Pattern:

NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX

Voluntary or Mandated:

Voluntary: Firm initiated