TriMed Inc.: Medical Device Recall in 2018 - (Recall #: Z-2262-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19
Class II
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Voluntary: Firm initiated