TriMed Inc.: Medical Device Recall in 2018 - (Recall #: Z-2262-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19

Product Classification:

Class II

Date Initiated: April 13, 2018
Date Posted: June 27, 2018
Recall Number: Z-2262-2018
Event ID: 80099
Reason for Recall:

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Status: Terminated
Product Quantity: 532
Code Information:

Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901

Distribution Pattern:

U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

Voluntary or Mandated:

Voluntary: Firm initiated