Product Description:
TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only, (01) 00842188101281 - Product Usage: The OHOOKL is a semi-tubular design bone plate with two offset intramedullary tines, which are intended to penetrate the small terminal fragment to provide additional stabilization.
Product Classification:
Class II
Date Initiated: August 15, 2018
Date Posted: November 20, 2019
Recall Number: Z-0366-2020
Event ID: 83984
Reason for Recall:
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully qualified.
Status: Terminated
Product Quantity: 583 Bone Plates
Code Information:
Lot # 110631,
Distribution Pattern:
US Nationwide distribution in the states of MD, UT.
Voluntary or Mandated:
Voluntary: Firm initiated