TriMed Inc.: Medical Device Recall in 2024 - (Recall #: Z-1791-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only

Product Classification:

Class II

Date Initiated: March 19, 2024
Date Posted: May 22, 2024
Recall Number: Z-1791-2024
Event ID: 94381
Reason for Recall:

Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.

Status: Ongoing
Product Quantity: 142 units
Code Information:

UDI: 00842188127878/ Lot # QM23086

Distribution Pattern:

US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.

Voluntary or Mandated:

Voluntary: Firm initiated