Product Description:

Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Product Classification:

Class II

Date Initiated: December 5, 2022
Date Posted: February 8, 2023
Recall Number: Z-1037-2023
Event ID: 91314
Reason for Recall:

Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.

Status: Ongoing
Product Quantity: 61 units
Code Information:

GTIN: 00859464007281 Lot Number: All Lot Numbers Serial Number: All Serial Numbers

Distribution Pattern:

US Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.

Voluntary or Mandated:

Voluntary: Firm initiated