Tyber Medical: Medical Device Recall in 2025 - (Recall #: Z-1209-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

A.L.P.S. mvX - ANATOMIC LAT FIB PLATE 4H LT: Anatomic Lateral Fibula Plate, 4-hole, Left Intended for Fixation of fractures of the distal tibia Model/Catalog Number: 770708041

Product Classification:

Class II

Date Initiated: January 21, 2025
Date Posted: March 5, 2025
Recall Number: Z-1209-2025
Event ID: 96212
Reason for Recall:

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Status: Ongoing
Product Quantity: 412 units
Code Information:

UDI-DI; 00196449015528 Lot Numbers 24043BU03, 267844, 24043BU02, 261286

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated