Tyber Medical: Medical Device Recall in 2025 - (Recall #: Z-1214-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H RT: Anatomic Lateral Fibula Plate, 8-hole, Right Indicated for use in Fixation of fractures of the distal tibia Model/Catalog Number: 77070802

Product Classification:

Class II

Date Initiated: January 21, 2025
Date Posted: March 5, 2025
Recall Number: Z-1214-2025
Event ID: 96212
Reason for Recall:

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Status: Ongoing
Product Quantity: 212 units
Code Information:

UDI-DI 00196449015573; Lot Numbers 265785, 24036BU01, 261291

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated