UFSK - International OSYS Gmbh: Medical Device Recall in 2023 - (Recall #: Z-2212-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006

Product Classification:

Class II

Date Initiated: May 31, 2023
Date Posted: July 26, 2023
Recall Number: Z-2212-2023
Event ID: 92534
Reason for Recall:

Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position

Status: Ongoing
Product Quantity: 239 chairs US
Code Information:

Art.no. / REF no.: 15000004 Serial number range: 2000- 2999 Art.no. / REF no.: 15000006 Serial number range: 3000-4506

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated