UIH Technologies LLC: Medical Device Recall in 2025 - (Recall #: Z-0220-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 780 MD, Rx Only
Product Classification:
Class II
Date Initiated: September 12, 2025
Date Posted: October 29, 2025
Recall Number: Z-0220-2026
Event ID: 97668
Reason for Recall:
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Status: Ongoing
Product Quantity: 2 units
Code Information:
UDI: (01)06971576831050/Serial Numbers: 684001, 684002
Distribution Pattern:
US: OUS:
Voluntary or Mandated:
Voluntary: Firm initiated