ulrich medical USA Inc: Medical Device Recall in 2020 - (Recall #: Z-1350-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
Product Classification:
Class II
Date Initiated: December 31, 2019
Date Posted: March 4, 2020
Recall Number: Z-1350-2020
Event ID: 84638
Reason for Recall:
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Status: Terminated
Product Quantity: 179 units
Code Information:
Lot Numbers: U010577, U016620, U019084
Distribution Pattern:
AZ, TX, GA, FL, AR, CA
Voluntary or Mandated:
Voluntary: Firm initiated