Product Description:

MultiLab Series II 2CP & 2CP Express non-invasive vascular diagnostic device, Model Number 11996-0000-01

Product Classification:

Class II

Date Initiated: February 10, 2023
Date Posted: March 8, 2023
Recall Number: Z-1193-2023
Event ID: 91564
Reason for Recall:

The current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism.

Status: Ongoing
Product Quantity: 1706 units
Code Information:

All Serial Numbers

Distribution Pattern:

Worldwide distribution - US Nationwide and Global distribution.

Voluntary or Mandated:

Voluntary: Firm initiated