Unomedical A/S: Medical Device Recall in 2023 - (Recall #: Z-0234-2024)
See the recall detail below. You can also see other recalls from the same firm in 2023.
VariSoft infusion set, Model Number 1002830
Class I
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
GTIN 05705244018792, Serial Numbers: 5388368; GTIN 05705244018785, Serial Numbers: 5388368
US Nationwide distribution in the state of CA.
Voluntary: Firm initiated