Varian Medical Systems, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0526-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Product Classification:

Class II

Date Initiated: April 30, 2014
Date Posted: December 24, 2014
Recall Number: Z-0526-2015
Event ID: 69637
Reason for Recall:

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

Status: Terminated
Product Quantity: 32 units
Code Information:

Part Number P1003290001

Distribution Pattern:

Worldwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated