Ventana Medical Systems Inc: Medical Device Recall in 2018 - (Recall #: Z-3010-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

VENTANA anti-Helicobacter Pylori (SP48) PAB 250, Catalog Number 06425623001, model 790-1015 Immunohistochemistry (IHC) for in vitro diagnostic use.

Product Classification:

Class I

Date Initiated: August 2, 2018
Date Posted: September 26, 2018
Recall Number: Z-3010-2018
Event ID: 80807
Reason for Recall:

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Status: Terminated
Product Quantity: 2753 units
Code Information:

Lot Numbers: Y12986, Y16448, Y24119, Y16912Z, UDI: 4015630970018

Distribution Pattern:

worldwide

Voluntary or Mandated:

Voluntary: Firm initiated