Ventana Medical Systems Inc: Medical Device Recall in 2022 - (Recall #: Z-0894-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
BenchMark ULTRA and DISCOVERY ULTRA Instruments
Class II
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
Serial Numbers Less-Than-Or-Equal-To 322030
US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
Voluntary: Firm initiated