Villa Sistemi Medicali S.P.A.: Medical Device Recall in 2017 - (Recall #: Z-1430-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
Apollo EZ; Model: 9784152036 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures
Product Classification:
Class II
Date Initiated: December 27, 2016
Date Posted: March 15, 2017
Recall Number: Z-1430-2017
Event ID: 76367
Reason for Recall:
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Status: Terminated
Product Quantity: 65 devices in total
Code Information:
Serial Number: 13101261
Distribution Pattern:
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
Voluntary or Mandated:
Voluntary: Firm initiated