VIRTUAL INCISION CORPORATION: Medical Device Recall in 2025 - (Recall #: Z-1905-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
Product Classification:
Class II
Date Initiated: April 23, 2025
Date Posted: June 11, 2025
Recall Number: Z-1905-2025
Event ID: 96831
Reason for Recall:
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
Status: Completed
Product Quantity: 3 units
Code Information:
UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213
Distribution Pattern:
US Distribution to States: CA, FL, NE
Voluntary or Mandated:
Voluntary: Firm initiated