Voco GmbH: Medical Device Recall in 2024 - (Recall #: Z-1142-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dental cement

Product Classification:

Class II

Date Initiated: January 25, 2024
Date Posted: February 28, 2024
Recall Number: Z-1142-2024
Event ID: 93877
Reason for Recall:

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Status: Ongoing
Product Quantity: 1530 units
Code Information:

a) 2540, UDI/DI +E22125401, Lot Numbers: 2338699, 2347154; b)2543, UDI/DI +E22125431, Lot Numbers: 2344046; c) 2544, UDI/DI +E22125441, Lot Numbers: 2342759, 2439537; d) 2545, UDI/DI +E22125451, Lot Numbers: 2348503; e) 2546, UDI/DI +E22125461, Lot Numbers: 2344045, 2349452; f) 2547, UDI/DI +E22125471, Lot Numbers: 2349453; g) 2549, UDI/DI +E22125491, Lot Numbers: 2342760, 2344016, 2344047, 2351233

Distribution Pattern:

Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.

Voluntary or Mandated:

Voluntary: Firm initiated