Vortex Surgical Inc.: Medical Device Recall in 2026 - (Recall #: Z-1244-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD
Class II
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
1. Flat Vitrectomy Lens Pouch UDI 810123482290 Box UDI 810123482306 Lots 2410036 2411020 2. Volk Magnifying Lens Pouch UDI 810123482825 Box UDI 810123482832 Lot 2411019
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Voluntary: Firm initiated