Vortex Surgical Inc.: Medical Device Recall in 2026 - (Recall #: Z-1252-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
Vortex Surgical I2 Injection Kit, VS0500
Class II
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Pouch UDI 810123480920 Box UDI 810123481330 Lot 2411012
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Voluntary: Firm initiated