Vyaire Medical: Medical Device Recall in 2018 - (Recall #: Z-1554-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
AirLife Resuscitation, Pediatric, with mask, 40inch oxygen reservoir tubing, pressure-relief valve Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.
Class I
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Part Number: 2K8008, Lots: 00010002841, 0001006156, 0001006158, 0001009054, 0001015412, 0001015413, 0001018698, 0001018699, 0001025674, 0001030572
Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.
Voluntary: Firm initiated