W & H DentalWerk Burmoos GMBH - Site 1: Medical Device Recall in 2018 - (Recall #: Z-1351-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Foot Controls Model: S-N1 and S-N2 Product Usage: Variable foot control for use with drive units operating mechanical instruments.

Product Classification:

Class II

Date Initiated: December 21, 2017
Date Posted: April 18, 2018
Recall Number: Z-1351-2018
Event ID: 79533
Reason for Recall:

During continuous postmarket surveillance activities, we have noticed a potential malfunction of individual foot controls of the type S-N1/S-N2. In case the error occurs, the foot control might not switch back to zero position when the pedal is released. Consequently, the motor keeps rotating. If this happens during automatic thread cutting or automatic screwing-in of implants, it may, in the worst case, lead to a serious injury of the patient.

Status: Terminated
Product Quantity:
Code Information:

REF: 05046200 05046200 05046211 05046210 06202400 06202400 30285000 30285001 Serial number "SN" from to (including) 22400 23429 E02201 E02485 E01011 E01033 E01035 E01071 24315 34040 E01358 E01658 01001 02318

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated