Waismed Ltd: Medical Device Recall in 2016 - (Recall #: Z-2680-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle into the bone marrow of a patient and allows connection of syringe or IV administration set. The NIO-A is intended for adults above 12 years old and the NIO-P is intended for pediatric patients 3 - 12 years of age.

Product Classification:

Class II

Date Initiated: August 18, 2016
Date Posted: September 7, 2016
Recall Number: Z-2680-2016
Event ID: 75008
Reason for Recall:

WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.

Status: Terminated
Product Quantity: 50
Code Information:

1620005

Distribution Pattern:

WI, TX, CA, LA and the following countries: UK, Spain, Poland, Israel, Czech Republic, France, Belgium, Denmark, Italy, Cyprus, Austria, Greece, Slovenia, Australia, Germany, Estonia, New Zealand, Switzerland, Costa Rica, Guatemala, Canada, Malaysia, Poland, Singapore, Lithuania, Romania

Voluntary or Mandated:

Voluntary: Firm initiated