WELCH ALLYN, INC/MORTARA: Medical Device Recall in 2020 - (Recall #: Z-1813-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
T12. Incorporates wireless electrocardiographic technology to achieve the real-time acquisition and RF transmission of simultaneous 12-lead ECG data with diagnostic quality while allowing the patient to be ambulatory.
Class II
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Model Numbers: T12-XXX-XXXXX, T12S-XXX-XXXXX.
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Voluntary: Firm initiated