WELCH ALLYN, INC/MORTARA: Medical Device Recall in 2020 - (Recall #: Z-1815-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.

Product Classification:

Class II

Date Initiated: March 18, 2020
Date Posted: May 6, 2020
Recall Number: Z-1815-2020
Event ID: 85297
Reason for Recall:

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Status: Terminated
Product Quantity: 78,521 units total
Code Information:

Model Numbers: S4-Q-ASX-XXX, S4-Q-AXX-XXX, S4-P-A, 9293-061-50, 9293-061-51, 9293-061-52, 9293-061-53.

Distribution Pattern:

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Voluntary or Mandated:

Voluntary: Firm initiated