WELCH ALLYN, INC/MORTARA: Medical Device Recall in 2020 - (Recall #: Z-1816-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications.
Class II
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Model Numbers: SUR12-XXX-XXXXX, SUR19-XXX-XXXXX, 9293-033-52, 9293-033-53.
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Voluntary: Firm initiated