WELCH ALLYN, INC/MORTARA: Medical Device Recall in 2020 - (Recall #: Z-1817-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
Class II
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Model Numbers: QS-6XX-XXXXX, XSCRIBE-5XX-XXXXX, XSCRIBE-6XX-XXXXX.
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Voluntary: Firm initiated