Welch Allyn Inc: Medical Device Recall in 2021 - (Recall #: Z-2482-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Braun ThermoScan PRO 6000 Ear Thermometer-indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment. Model: 6000
Class II
If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient
All Devices in distribution. UDI: 00732094309027 DDDYYKXXXXX DDD = 3 DIGIT JULIAN DAY (EXAMPLE - 234). YY = LAST 2 DIGITS OF JULIAN YEAR (EXAMPLE - 13) K = KEYTRONIC XXXXX = 5 DIGIT BASE 10 SERIALIZED NUMBER.
Worldwide distribution - US Nationwide distribution and the countries of Afghanistan, Andorra, Australia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Canada, Chile, China, Colombia, Congo, Costa Rica, C¿te D'ivoire (Ivory Coast), Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Estonia, Finland, France, Frenc.Polynesia, French, Guiana, Germany, Greece, Guadeloupe, Guam, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Monaco, Myanmar, N.Mariana Islnd, Netherlands (the), New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Utd.Arab Emir., Vietnam, Zimbabwe.
Voluntary: Firm initiated