West Pharmaceutical Services AZ, Inc.: Medical Device Recall in 2024 - (Recall #: Z-0283-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Brand Name: Vial2Bag Advanced¿ Product Name: Vial2Bag Advanced¿ 20mm Admixture Device Model/Catalog Number: 6070030 Software Version: Not applicable Product Description: The Vial2Bag Advanced 20mm Admixture Device is a single use, fluid transfer device that allows for the reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is provided as a sterile, non-pyrogenic product. The device is intended to be used with standard drug vials with a seal diameter of 20mm and an elastomeric stopper. The Vial2Bag Advanced 20mm Admixture Device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV administration set. The device does not contain any medicinal substances and there are no additional accessories needed or provided with the Vial2Bag Advanced 20mm Admixture Device for the device to meet its intended purpose. Component: No

Product Classification:

Class II

Date Initiated: September 24, 2024
Date Posted: November 6, 2024
Recall Number: Z-0283-2025
Event ID: 95449
Reason for Recall:

Due to packaging damage from heat exposure during shipping/transportation that may result in a breach in the sterile barrier.

Status: Ongoing
Product Quantity: 129600
Code Information:

Lot Code: Recall Product 1: Product Code 6080030, Lot# K188, UDI (01)10850022888011(17)270430(10)K188\F(90)36098150\F(91)6070030 Recall Product 2: Product Code 6080030, Lot# K194, UDI (01)10850022888011(17)270430(10)K194\F(90)36098150\F(91)6070030

Distribution Pattern:

A U.S. Distributor, Progressive Medical Inc. (PMI) currently supplies the Vial2Bag Advanced 20mm Admixture Device to hospitals in the U.S. (end user). PMI distributed the impacted Vial2Bag Advanced 20mm Admixture Device lots to the end users. U.S.: Distributor located in MO U.S. Customers: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NE, NM, NV, NY, OH, OK, OR, PA,SC, TN, TX, UT, WA, and WI.

Voluntary or Mandated:

Voluntary: Firm initiated