Western/Scott Fetzer Company: Medical Device Recall in 2021 - (Recall #: Z-0269-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Western Oxytote DTE Digital VIPR System (with D-size cylinder) Model: MNDS-602
Product Classification:
Class II
Date Initiated: October 16, 2021
Date Posted: December 1, 2021
Recall Number: Z-0269-2022
Event ID: 88877
Reason for Recall:
VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
Status: Ongoing
Product Quantity: 40 units
Code Information:
Units distributed from 05/29/20 to 08/31/21. Digital VIPR units manufactured and distributed after 08/31/21 are outside the scope of this recall and will be recognized by a green and silver control knob sticker (green logo, silver background)
Distribution Pattern:
CT, MD, MT, OH
Voluntary or Mandated:
Voluntary: Firm initiated