Westmed, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2434-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.
Class II
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
101118T42, 101118T44, 102918T01, 080618T02, 080618T03 , 101118T43, 102918T02.
Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.
Voluntary: Firm initiated