Westmed, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0742-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Westmed DISPOSABLE RESPIRATORY PRODUCTS NON-REBREATHER MASK, REF 3228-e, RX Only, CE0482, (01)10709078008020 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.

Product Classification:

Class II

Date Initiated: November 11, 2020
Date Posted: January 6, 2021
Recall Number: Z-0742-2021
Event ID: 86903
Reason for Recall:

There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.

Status: Terminated
Product Quantity: 37,300 units
Code Information:

Lot # 081120N42, 090220N29, 091020N88

Distribution Pattern:

Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated