Wilson-Cook Medical Inc.: Medical Device Recall in 2023 - (Recall #: Z-1979-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-10, REF G21049; b) HEMO-10-EU, REF G21346
Product Classification:
Class II
Date Initiated: May 22, 2023
Date Posted: June 28, 2023
Recall Number: Z-1979-2023
Event ID: 92345
Reason for Recall:
There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.
Status: Ongoing
Product Quantity: 14,877 units
Code Information:
UDI/DI : a) 00827002210493, b) 00827002213463; All unexpired lot numbers
Distribution Pattern:
US Nationwide - Worldwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated