Wilson-Cook Medical Inc.: Medical Device Recall in 2024 - (Recall #: Z-0582-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Product Classification:
Class II
Date Initiated: October 31, 2024
Date Posted: December 11, 2024
Recall Number: Z-0582-2025
Event ID: 95755
Reason for Recall:
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
Status: Ongoing
Product Quantity: 46 devices
Code Information:
UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.
Distribution Pattern:
Distribution was OUS only to Argentina, Brazil, and Israel.
Voluntary or Mandated:
Voluntary: Firm initiated