Wright Medical Technology, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1490-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

CROSSCHECK(R) Plating system, MTP NX Plate, Left, Recon, REF CCP-MPN2L. Orthopedic use.

Product Classification:

Class II

Date Initiated: March 6, 2015
Date Posted: April 29, 2015
Recall Number: Z-1490-2015
Event ID: 70859
Reason for Recall:

The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction

Status: Terminated
Product Quantity: 167 units
Code Information:

Lot Numbers: 3504917, 35049-17A, 35049-17A-1, 35049-18, 362791, 36279-1-1

Distribution Pattern:

Nationwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated