Wright Medical Technology, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1499-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

CROSSCHECK(R) Plating system, Utility Plate, 7-Hole, REF CCP-UTN7. Orthopedic use.

Product Classification:

Class II

Date Initiated: March 6, 2015
Date Posted: April 29, 2015
Recall Number: Z-1499-2015
Event ID: 70859
Reason for Recall:

The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.

Status: Terminated
Product Quantity: 151 units
Code Information:

Lot Numbers: 3504937, 35049-37A, 35049-37A-1, 35049-38, 362806

Distribution Pattern:

Nationwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated