Wright Medical Technology, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1366-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Product Classification:

Class II

Date Initiated: June 14, 2022
Date Posted: July 20, 2022
Recall Number: Z-1366-2022
Event ID: 90399
Reason for Recall:

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

Status: Ongoing
Product Quantity: 231
Code Information:

Lot codes: 955401, 1168853, 1349302, 1956936, 2486857

Distribution Pattern:

Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico

Voluntary or Mandated:

Voluntary: Firm initiated