Wright Medical Technology, Inc.: Medical Device Recall in 2023 - (Recall #: Z-2105-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.
Product Classification:
Class II
Date Initiated: June 13, 2023
Date Posted: July 12, 2023
Recall Number: Z-2105-2023
Event ID: 92460
Reason for Recall:
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
Status: Ongoing
Product Quantity: 106 devices
Code Information:
Lot numbers 2656815 and 2762126, UDI-DI 00889797003926.
Distribution Pattern:
There was no U.S. distribution, including government and military. Foreign distribution was made to Australia, Belgium, Canada, Chile, China, Colombia, France, Germany, Italy, Mexico, Netherlands, Spain, Sweden, Switzerland, Taiwan, and the United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated