Xcision Medical Systems, LLC: Medical Device Recall in 2024 - (Recall #: Z-0750-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.
Class II
Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected
00850002615029(11)230321(21)TPS0018 (Distributed in the US) 00850002615029(11)230629(21)TPS0013 (Distributed in the US) 00850002615005(11)21H24(21)TPS0016 (Distributed in the US) 00850002615029(11)210629(21 )TPS0014 (Italy)
Worldwide - US Nationwide distribution in the states of MD, PA, TX and the country of Italy.
Voluntary: Firm initiated