Xeridiem Mediem Medical Devices Inc: Medical Device Recall in 2016 - (Recall #: Z-0968-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
EndoVive 3s Low Profile Balloon Kits Part Number: M00548420 (XMD P/N 70-0050-914)
Product Classification:
Class II
Date Initiated: December 23, 2015
Date Posted: March 2, 2016
Recall Number: Z-0968-2016
Event ID: 73024
Reason for Recall:
A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.
Status: Terminated
Product Quantity: 40
Code Information:
Part Number: M00548420 (XMD P/N 70-0050-914) Lot Number: 929610
Distribution Pattern:
Nationwide distribution to MA only.
Voluntary or Mandated:
Voluntary: Firm initiated