Zeiss, Carl Inc: Medical Device Recall in 2018 - (Recall #: Z-2441-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Product Classification:
Class II
Date Initiated: May 29, 2018
Date Posted: July 25, 2018
Recall Number: Z-2441-2018
Event ID: 80418
Reason for Recall:
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Status: Terminated
Product Quantity: 8
Code Information:
432332-9020-000
Distribution Pattern:
US Nationwide Distribution
Voluntary or Mandated:
FDA Mandated