Zimmer Biomet, Inc.: Medical Device Recall in 2017 - (Recall #: Z-3154-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
NCB Periprosthetic Plate System Drill Bit With Quick-Coupling 3.5 mm Diameter 195 mm Length, Item Number/EDI 0200024325, Nonsterile. orthopedic surgical instrument.
Class II
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Lot Numbers: 63385956, 63405774 & 63445667
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
Voluntary: Firm initiated