Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-0424-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Affixus Hip Fracture Nail, LH 125 DEG 11MM X 320MM

Product Classification:

Class II

Date Initiated: August 7, 2017
Date Posted: February 7, 2018
Recall Number: Z-0424-2018
Event ID: 78763
Reason for Recall:

During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.

Status: Terminated
Product Quantity: 638 units
Code Information:

Item No. 814411320; Lot No. (Exp. Date) UDI: 459230 (05/19/2027) UDI (01)00887868038464 (17)270519 (10)459230, 701830 (06/03/2027) UDI (01)00887868038464 (17)270603 (10)701830, 893480 (06/12/2027) UDI (01) 00887868038464 (17)270612 (10)893480

Distribution Pattern:

US nationwide distribution. Also Portugal and the Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated