Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-0721-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Zimmer Segmental System, ZSS Distal Femur. Sizes as follows: a. Size B-LT (Item No. 00585001201); b. Size B-RT (Item No. 00585001202); c. Size C-LT (Item No. 00585001301); d. Size C-RT (Item No. 00585001302) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.

Product Classification:

Class II

Date Initiated: August 1, 2017
Date Posted: March 7, 2018
Recall Number: Z-0721-2018
Event ID: 78762
Reason for Recall:

The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.

Status: Terminated
Product Quantity: 3,350 units total
Code Information:

a. Item No. 00585001201 (Lot Expiration Date Before 07/31/2026); b. Item No. 00585001202 (Lot Expiration Date Before 07/31/2026); c. Item No. 00585001301 (Lot Expiration Date Before 08/31/2026); d. Item No. 00585001302 (Lot Expiration Date Before 07/31/2026)

Distribution Pattern:

Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico; and to countries of: Austria, Australia, Bahamas, Brazil, Canada, China, Dominican Republic, England, France, Germany, Israel, Japan, South Korea, Lebanon, Malaysia, Mexico, Netherlands, Panama, Singapore, Spain, Switzerland, and Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated