Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1024-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
VERSYS 6 INCH BEADED FC 14X160MM STD BODY EXT & STD NECK Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Class II
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Item Number: 00-7843-014-26 Lot Number: 61741806 Item Number: 00-7843-014-06 Lot Numbers: 61886028 61752736 62120676
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
Voluntary: Firm initiated