Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1028-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
VERSYS 7.5 INCH BEADED FC REV 13.5X190MM STRAIGHT Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Class II
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Item Number: 00-7843-013-18 Lot Number: 62158840
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
Voluntary: Firm initiated