Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1071-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
TunneLoc Tibial Fixation Device Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.
Product Classification:
Class II
Date Initiated: October 11, 2017
Date Posted: March 21, 2018
Recall Number: Z-1071-2018
Event ID: 79263
Reason for Recall:
Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.
Status: Terminated
Product Quantity: 14,461
Code Information:
906512, 906513, 906514, 906515
Distribution Pattern:
US Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated