Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2240-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

3.2MM X 510MM COCR THD TIP; 14-443053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Product Classification:

Class II

Date Initiated: May 16, 2018
Date Posted: June 27, 2018
Recall Number: Z-2240-2018
Event ID: 80193
Reason for Recall:

The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.

Status: Terminated
Product Quantity: 304
Code Information:

001390 105790 324290 974210 820900 512140 428260 915580 017930 671970 055480 615280 506500 569970 053580 367190 932620 015010 141060 820240 941850 682590 401830

Distribution Pattern:

Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.

Voluntary or Mandated:

Voluntary: Firm initiated